Skip to content
Fencer Intel
medium 4.0

CVE-2011-3386 json

Unspecified vulnerability in Medtronic Paradigm wireless insulin pump 512, 522, 712, and 722 allows remote attackers to modify the delivery…

published 2011-09-02 15 y ago modified 2026-06-16 4 mo ago also EUVD-2011-3349

cvss
4.0
version 2.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
1.5%
EPSS 1.5% chance of exploitation in 30 days, percentile 73%
public exploits
0
none known

Summary

Unspecified vulnerability in Medtronic Paradigm wireless insulin pump 512, 522, 712, and 722 allows remote attackers to modify the delivery of an insulin bolus dose and cause a denial of service (adverse human health effects) via unspecified vectors involving wireless communications and knowledge of the device's serial number, as demonstrated by Jerome Radcliffe at the Black Hat USA conference in August 2011. NOTE: the vendor has disputed the severity of this issue, saying "we believe the risk of deliberate, malicious, or unauthorized manipulation of medical devices is extremely low... we strongly believe it would be extremely difficult for a third-party to wirelessly tamper with your insulin pump... you would be able to detect tones on the insulin pump that weren't intentionally programmed and could intervene accordingly."

Exploit prediction

1.45% chance of exploitation in the next 30 days, higher than 73% of all scored CVEs (FIRST EPSS, 2026-11-09).

CVSS scores

assigned by version score vector
NVD 2.0 4.0 AV:N/AC:H/Au:N/C:N/I:P/A:P

References

Sources

Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.