Skip to content
Fencer Intel
medium 6.8

CVE-2014-5431 json

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded pas…

published 7y ago modified 4mo ago also EUVD-2014-5319

cvss
6.8
version 3.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.4%
EPSS 0.4% chance of exploitation in 30 days, percentile 31%
public exploits
0
none known

Summary

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Is CVE-2014-5431 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Exploit prediction

0.38% chance of exploitation in the next 30 days, higher than 31% of all scored CVEs (FIRST EPSS, 2026-11-09).

CVSS scores

assigned by version score vector
NVD 3.0 6.8 CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NVD 2.0 4.6 AV:L/AC:L/Au:N/C:P/I:P/A:P

Affected products

  • Baxter SIGMA Spectrum Infusion System

Classification

Weakness
  • CWE-259 Use of Hard-coded Password
  • CWE-798 Use of Hard-coded Credentials

References

Sources

Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.