Skip to content
Fencer Intel
critical 9.8

CVE-2014-5432 json

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible vi…

published 7y ago modified 4mo ago also EUVD-2014-5320

cvss
9.8
version 3.0, NVD
cisa kev
Not listed
not in CISA's catalog
epss
2.6%
EPSS 2.6% chance of exploitation in 30 days, percentile 85%
public exploits
0
none known

Summary

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Is CVE-2014-5432 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Exploit prediction

2.57% chance of exploitation in the next 30 days, higher than 85% of all scored CVEs (FIRST EPSS, 2026-11-09).

CVSS scores

assigned by version score vector
NVD 3.0 9.8 CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NVD 2.0 7.5 AV:N/AC:L/Au:N/C:P/I:P/A:P

Affected products

  • Baxter SIGMA Spectrum Infusion System

Classification

Weakness
  • CWE-592 DEPRECATED: Authentication Bypass Issues
  • CWE-287 Improper Authentication

References

Sources

Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.