CVE-2015-1012 json
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System.
published 2019-03-25 7 y ago modified 2026-06-17 4 mo ago also EUVD-2015-1156
- cvss
- 7.5
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.8%
- EPSS 0.8% chance of exploitation in 30 days, percentile 55%
- public exploits
- 0
- none known
Summary
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSA-15-132-01 updated cvss 6.4
Hospira LifeCare PCA Infusion System Vulnerabilities
-
Hospira LifeCare PCA Infusion System <=5.0affected: mitigation
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
cisa ICSA-15-125-01B updated cvss 6.4
Hospira LifeCare PCA Infusion System Vulnerabilities
-
Hospira LifeCare PCA Infusion System <=5.0affected: mitigation “For additional information about Hospira’s upcoming release, contact Hospira’s technical support at 1-800-241-4002.”
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.77% chance of exploitation in the next 30 days, higher than 55% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 7.5 | CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 5.0 | AV:N/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- Hospira LifeCare PCA Infusion System
Classification
References
-
ics-cert.us-cert.gov
/advisories/ICSA-15-125-01
mitigation, third-party-advisory, us-government-resource
Sources
- CVE List CVE-2015-1012
- NVD CVE-2015-1012
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.