CVE-2017-12716 json
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communi…
published 8y ago modified 4mo ago also EUVD-2017-4255
- cvss
- 6.5
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 15%
- public exploits
- 0
- none known
Summary
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Patch / fix
- abbott accent_firmware f0b.0e.7e
- abbott anthem_firmware f0b.0e.7e
- abbott accent_mri_firmware f10.08.6c
- abbott accent_st_firmware f10.08.6c
Is CVE-2017-12716 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-17-241-01 updated cvss 3.1
ICSMA-17-241-01_Abbott Laboratories ' Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI Pacemaker Vulnerabilities
-
Abbott Laboratories Accent MRI < august 28 Abbott Laboratories Accent/Anthem < august 28 Abbott Laboratories Assurity MRI < august 28 Abbott Laboratories Assurity/Allure < august 28affected: mitigation
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.24% chance of exploitation in the next 30 days, higher than 15% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 6.5 | CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- Abbott Laboratories Accent and Anthem
Classification
References
-
ics-cert.us-cert.gov
/advisories/ICSMA-17-241-01
third-party-advisory, us-government-resource
-
securityfocus.com
/bid/100523
vdb-entry, third-party-advisory
Sources
- CVE List CVE-2017-12716
- NVD CVE-2017-12716
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.