CVE-2017-9657 json
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the centr…
published 2018-04-30 8 y ago modified 2026-06-17 4 mo ago also EUVD-2017-18588
- cvss
- 6.5
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.8%
- EPSS 0.8% chance of exploitation in 30 days, percentile 55%
- public exploits
- 0
- none known
Summary
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Patch / fix
- philips intellivue_mx40_firmware b.06.18
Vendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-17-255-01 updated cvss 6.5
ICSMA-17-255-01_Philips' IntelliView MX40 Patient Worn Monitor (WLAN) Vulnerabilities
-
Philips IntelliVue MX40 Patient Worn Monitor (WLAN only) < B.06.18affected: mitigation usa.philips.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.77% chance of exploitation in the next 30 days, higher than 55% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 6.5 | CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:N/I:N/A:P |
Affected products
- Philips IntelliVue MX40 Patient Worn Monitor
Classification
References
-
usa.philips.com
/healthcare/about/customer-support/product-security
vendor-advisory
-
securityfocus.com
/bid/100813
vdb-entry, third-party-advisory
-
ics-cert.us-cert.gov
/advisories/ICSMA-17-255-01
third-party-advisory, us-government-resource