Skip to content
Fencer Intel
high 8.2

CVE-2018-10601 json

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550)…

published 8y ago modified 4mo ago also EUVD-2018-2673

cvss
8.2
version 3.1, NVD
cisa kev
Not listed
not in CISA's catalog
epss
0.4%
EPSS 0.4% chance of exploitation in 30 days, percentile 31%
public exploits
0
none known

Summary

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Is CVE-2018-10601 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-18-156-01 updated cvss 8.2

Philips' IntelliVue Patient and Avalon Fetal Monitors

  • Philips Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 F.0.G.0, J.3 Philips IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M Philips IntelliVue Patient Monitors MX (MX400-550) Rev J-M, (X3/MX100 for Rev M only)
    affected: vendor fix philips.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

Exploit prediction

0.37% chance of exploitation in the next 30 days, higher than 31% of all scored CVEs (FIRST EPSS, 2026-11-09).

CVSS scores

assigned by version score vector
NVD 3.1 8.2 CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H
NVD 2.0 5.4 AV:A/AC:M/Au:N/C:P/I:P/A:P

Affected products

  • ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors

Classification

Weakness

References

Sources

Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.