CVE-2018-18562 json
An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Uni…
published 7y ago modified 4mo ago also EUVD-2018-10282
- cvss
- 8.8
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.7%
- EPSS 0.7% chance of exploitation in 30 days, percentile 54%
- public exploits
- 0
- none known
Summary
An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface.
Patch / fix
- roche accu-chek_inform_ii_firmware 03.01.04
- roche cobas_h_232_firmware 03.01.04
- roche coaguchek_firmware 03.01.04
- roche base_unit_hub_firmware 03.01.04
Is CVE-2018-18562 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-18-310-01 updated cvss 8.0
Roche Diagnostics Point of Care Handheld Medical Devices (Update A)
-
Roche Diagnostics Point of Care handheld medical devices Accu-Chek Inform II, CoaguChek Pro II, CoaguChek XS Plus, CoaguChek XS Proaffected: vendor fix roche.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.72% chance of exploitation in the next 30 days, higher than 54% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 8.8 | CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:P/I:N/A:N |
Classification
- Weakness
-
- CWE-521 Weak Password Requirements
References
-
ics-cert.us-cert.gov
/advisories/ICSMA-18-310-01
mitigation, third-party-advisory, us-government-resource
-
securityfocus.com
/bid/105843
vdb-entry, third-party-advisory
Sources
- CVE List CVE-2018-18562
- NVD CVE-2018-18562