CVE-2018-8861 json
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 an…
published 8y ago modified 4mo ago also EUVD-2018-20469
- cvss
- 8.7
- version 3.0, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.4%
- EPSS 0.4% chance of exploitation in 30 days, percentile 34%
- public exploits
- 0
- none known
Summary
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.
Is CVE-2018-8861 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-18-123-01 updated cvss 6.1
Philips Brilliance Computed Tomography (CT) System (Update A)
-
Philips Brilliance 64 <=2.6.2 Philips Brilliance CT Big Bore <=2.3.5 Philips Brilliance iCT <=4.1.6, <=3.2.4affected: vendor fix incenter.medical.philips.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.40% chance of exploitation in the next 30 days, higher than 34% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.0 | 8.7 | CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:L |
| NVD | 2.0 | 6.8 | AV:L/AC:L/Au:N/C:C/I:C/A:P |
Affected products
- Philips Brilliance CT Scanners
Classification
- Weakness
-
- CWE-668 Exposure of Resource to Wrong Sphere
References
-
usa.philips.com
/healthcare/about/customer-support/product-security
vendor-advisory
-
securityfocus.com
/bid/104088
vdb-entry, third-party-advisory
-
ics-cert.us-cert.gov
/advisories/ICSMA-18-123-01
third-party-advisory, us-government-resource