CVE-2018-8868 json
Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
published 8y ago modified 4mo ago also EUVD-2018-20476
- cvss
- 6.2
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.3%
- EPSS 0.3% chance of exploitation in 30 days, percentile 27%
- public exploits
- 0
- none known
Summary
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Is CVE-2018-8868 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-18-179-01 updated cvss 6.2
Medtronic MyCareLink Patient Monitor
-
Medtronic 24950 MyCareLink Monitor Medtronic 24952 MyCareLink Monitoraffected: vendor fix medtronic.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.34% chance of exploitation in the next 30 days, higher than 27% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 6.2 | CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L |
| NVD | 3.0 | 6.4 | CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 6.9 | AV:L/AC:M/Au:N/C:C/I:C/A:C |
Affected products
- Medtronic 24950 MyCareLink Monitor
- Medtronic 24952 MyCareLink Monitor
Classification
- Weakness
-
- CWE-749 Exposed Dangerous Method or Function
References
- global.medtronic.com /xg-en/product-security/security-bulletins/mycarelink-6-28-18.html
-
ics-cert.us-cert.gov
/advisories/ICSMA-18-179-01
third-party-advisory, us-government-resource
Sources
- CVE List CVE-2018-8868
- NVD CVE-2018-8868
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.