CVE-2019-25722 json
Dräger SC Monitoring Devices Hard-coded Credentials and DoS
published 5mo ago modified 3mo ago also EUVD-2019-20158
- cvss
- 7.6
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 9%
- public exploits
- 0
- none known
Summary
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Is CVE-2019-25722 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerCISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- 5mo ago
Exploit prediction
0.19% chance of exploitation in the next 30 days, higher than 9% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 4.0 | 7.2 | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N |
| CNA | 3.1 | 7.6 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| CNA | 4.0 | 7.2 | CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Dräger SC 6002XL
- Dräger SC6802XL
- Dräger SC 7000
- Dräger SC8000
- Dräger SC90000 XL
Classification
- Weakness
-
- CWE-798 Use of Hard-coded Credentials
References
Sources
- CVE List CVE-2019-25722
- NVD CVE-2019-25722
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.