CVE-2019-6540 json
Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
published 7y ago modified 4mo ago also EUVD-2019-16099
- cvss
- 6.5
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 9%
- public exploits
- 0
- none known
Summary
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Is CVE-2019-6540 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-19-080-01 updated cvss 6.5
Medtronic Conexus Radio Frequency Telemetry Protocol (Update C)
-
Medtronic Amplia CRT-D Medtronic Brava CRT-D Medtronic CareLink 2090 Programmer Medtronic CareLink Monitor 2490C Medtronic Claria CRT-D Medtronic Compia CRT-Daffected: vendor fix medtronic.com
+16 more productsfewer
Medtronic Concerto CRT-D Medtronic Concerto II CRT-D Medtronic Consulta CRT-D Medtronic Evera ICD Medtronic Maximo II CRT-D and ICD Medtronic Mirro ICD Medtronic Mirro MRI ICD Medtronic MyCareLink Monitor 24950, 24952 Medtronic Nayamed ND ICD Medtronic Primo ICD Medtronic Protecta ICD and CRT-D Medtronic Secura ICD Medtronic Virtuoso ICD Medtronic Virtuoso II ICD Medtronic Visia AF ICD Medtronic Viva CRT-D
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.20% chance of exploitation in the next 30 days, higher than 9% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:P/I:N/A:N |
Affected products
- Medtronic Conexus Radio Frequency Telemetry Protocol
- Medtronic MyCareLink Monitor
- Medtronic CareLink Monitor
- Medtronic CareLink 2090 Programmer
- Medtronic Amplia CRT-D
- Medtronic Claria CRT-D
- Medtronic Compia CRT-D
- Medtronic Concerto CRT-D
- Medtronic Concerto II CRT-D
- Medtronic Consulta CRT-D
- Medtronic Evera ICD
- Medtronic Maximo II CRT-D
- Medtronic Maximo II ICD
- Medtronic Mirro ICD
- Medtronic Nayamed ND ICD
- Medtronic Primo ICD
- Medtronic Protecta ICD, Protecta CRT-D
- Medtronic Secura ICD
- Medtronic Virtuoso ICD
- Medtronic Virtuoso II ICD
- Medtronic Visia AF ICD
- Medtronic Viva CRT-D
- Medtronic Brava CRT-D
- Medtronic Mirro MRI ICD
Classification
- Weakness
-
- CWE-319 Cleartext Transmission of Sensitive Information
References
-
ics-cert.us-cert.gov
/advisories/ICSMA-19-080-01
third-party-advisory, us-government-resource
-
securityfocus.com
/bid/107544
vdb-entry, broken-link
Sources
- CVE List CVE-2019-6540
- NVD CVE-2019-6540
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.