CVE-2020-16216 json
Philips Patient Monitoring Devices Improper Input Validation
published 6y ago modified 4mo ago also EUVD-2020-8182
- cvss
- 6.5
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.7%
- EPSS 0.7% chance of exploitation in 30 days, percentile 53%
- public exploits
- 0
- none known
Summary
In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart.
Is CVE-2020-16216 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-254-01 updated cvss 6.5
Philips Patient Monitoring Devices (Update C)
-
Philips IntelliVue patient monitor MP2-MP90 <=N Philips IntelliVue patient monitor MX100 <=N Philips IntelliVue patient monitor MX400-MX850 <=Naffected: vendor fix, mitigation “Philips released the following versions to remediate reported vulnerabilities:” usa.philips.com
-
Philips IntelliVue X2 <=N Philips IntelliVue X3 <=Naffected: mitigation “Philips released the following versions to remediate reported vulnerabilities:” incenter.medical.philips.com
-
Philips Patient Information Center iX (PICiX) B.02, C.02, C.03 Philips PerformanceBridge Focal Point A.01affected: vendor fix, mitigation “Philips released the following versions to remediate reported vulnerabilities:” incenter.medical.philips.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.70% chance of exploitation in the next 30 days, higher than 53% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 6.1 | AV:A/AC:L/Au:N/C:N/I:N/A:C |
Affected products
- Philips IntelliVue patient monitors
- Philips IntelliVue
Classification
- Weakness
-
- CWE-20 Improper Input Validation
References
-
us-cert.cisa.gov
/ics/advisories/icsma-20-254-01
third-party-advisory, us-government-resource
- philips.com /productsecurity
Sources
- CVE List CVE-2020-16216
- NVD CVE-2020-16216
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.