CVE-2020-16224 json
Philips Patient Monitoring Devices Improper Handling of Length Parameter Inconsistency
published 6y ago modified 4mo ago also EUVD-2020-8190
- cvss
- 6.5
- version 3.1, NVD
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.6%
- EPSS 0.6% chance of exploitation in 30 days, percentile 48%
- public exploits
- 0
- none known
Summary
In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
Is CVE-2020-16224 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-254-01 updated cvss 6.5
Philips Patient Monitoring Devices (Update C)
-
Philips IntelliVue patient monitor MP2-MP90 <=N Philips IntelliVue patient monitor MX100 <=N Philips IntelliVue patient monitor MX400-MX850 <=Naffected: vendor fix, mitigation “Philips released the following versions to remediate reported vulnerabilities:” usa.philips.com
-
Philips IntelliVue X2 <=N Philips IntelliVue X3 <=Naffected: mitigation “Philips released the following versions to remediate reported vulnerabilities:” incenter.medical.philips.com
-
Philips Patient Information Center iX (PICiX) B.02, C.02, C.03 Philips PerformanceBridge Focal Point A.01affected: vendor fix, mitigation “Philips released the following versions to remediate reported vulnerabilities:” incenter.medical.philips.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
Exploit prediction
0.59% chance of exploitation in the next 30 days, higher than 48% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| NVD | 3.1 | 6.5 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| NVD | 2.0 | 3.3 | AV:A/AC:L/Au:N/C:N/I:N/A:P |
Affected products
- Philips Patient Information Center iX (PICiX)
Classification
- Weakness
-
- CWE-130 Improper Handling of Length Parameter Inconsistency
References
-
us-cert.cisa.gov
/ics/advisories/icsma-20-254-01
third-party-advisory, us-government-resource
- philips.com /productsecurity
Sources
- CVE List CVE-2020-16224
- NVD CVE-2020-16224
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.