CVE-2020-25150 json
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
published 4y ago modified 4mo ago also EUVD-2020-17841
- cvss
- 7.6
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 1.5%
- EPSS 1.5% chance of exploitation in 30 days, percentile 74%
- public exploits
- 0
- none known
Summary
A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands.
Is CVE-2020-25150 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-296-02 updated cvss 7.6
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
-
B. Braun Medical Battery pack with Wi-Fi <= U61, <= L81 B. Braun Medical Data module compactplus A10, A11 B. Braun Medical SpaceCom <= U61, <= L81affected: vendor fix, mitigation “Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).” bbraun.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- last yr.
Exploit prediction
1.53% chance of exploitation in the next 30 days, higher than 74% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.6 | CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L |
| NVD | 3.1 | 8.8 | CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H |
| NVD | 2.0 | 9.0 | AV:N/AC:L/Au:S/C:C/I:C/A:C |
Affected products
- B. Braun Melsungen AG SpaceCom
- B. Braun Melsungen AG Battery pack with Wi-Fi
- B. Braun Melsungen AG Data module compactplus
Classification
References
-
cisa.gov
/uscert/ics/advisories/icsma-20-296-02
third-party-advisory, us-government-resource
-
bbraun.com
/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
broken-link
Sources
- CVE List CVE-2020-25150
- NVD CVE-2020-25150
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.