CVE-2020-25166 json
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
published 4y ago modified 4mo ago also EUVD-2020-17857
- cvss
- 7.6
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 41%
- public exploits
- 0
- none known
Summary
An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.
Is CVE-2020-25166 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-20-296-02 updated cvss 6.8
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
-
B. Braun Medical Battery pack with Wi-Fi <= U61, <= L81 B. Braun Medical Data module compactplus A10, A11 B. Braun Medical SpaceCom <= U61, <= L81affected: vendor fix, mitigation “Wireless networks should be implemented using industry standard encryption and should be equipped with intrusion detection systems (IDS) and/or intrusion prevention systems (IPS).” bbraun.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- partial
- Assessed
- last yr.
Exploit prediction
0.47% chance of exploitation in the next 30 days, higher than 41% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 7.6 | CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L |
| NVD | 3.1 | 7.1 | CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L |
| NVD | 2.0 | 7.5 | AV:N/AC:L/Au:S/C:N/I:C/A:P |
Affected products
- B. Braun Melsungen AG SpaceCom
- B. Braun Melsungen AG Battery pack with Wi-Fi
- B. Braun Melsungen AG Data module compactplus
Classification
- Weakness
-
- CWE-347 Improper Verification of Cryptographic Signature
References
-
cisa.gov
/uscert/ics/advisories/icsma-20-296-02
third-party-advisory, us-government-resource
-
bbraun.com
/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
broken-link
Sources
- CVE List CVE-2020-25166
- NVD CVE-2020-25166
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.