Skip to content
Fencer Intel
medium 5.3

CVE-2021-23195 json

Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing

published 4y ago modified 4mo ago also EUVD-2021-10301

cvss
5.3
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.8%
EPSS 0.8% chance of exploitation in 30 days, percentile 58%
public exploits
0
none known

Summary

Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server.

Patch / fix

  • Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3

Is CVE-2021-23195 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-21-355-01 updated cvss 5.3

Fresenius Kabi Agilia Connect Infusion System (Update A)

  • Fresenius Kabi Agilia Connect WiFi module of the pumps <= D25 Fresenius Kabi Agilia Link+ <= 3.0 D15 Fresenius Kabi Agilia Partner maintenance software <= 3.3.0
    affected: mitigation “Fresenius Kabi has created new versions to address these vulnerabilities: Link+ v3.0 (D16 or later)” fresenius-kabi.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
yes
Technical impact
partial
Assessed
last yr.

Exploit prediction

0.85% chance of exploitation in the next 30 days, higher than 58% of all scored CVEs (FIRST EPSS, 2026-11-09).

CVSS scores

assigned by version score vector
CNA 3.1 5.3 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
NVD 3.1 5.3 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
NVD 2.0 5.0 AV:N/AC:L/Au:N/C:P/I:N/A:N

Affected products

  • Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard)

Classification

Weakness
  • CWE-548 Exposure of Information Through Directory Listing
  • CWE-200 Exposure of Sensitive Information to an Unauthorized Actor

References

Sources

Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.