CVE-2024-48966 json
Life2000 service tools for test and calibration do not support user authentication
published last yr. modified 4mo ago also EUVD-2024-43153
- cvss
- 10.0
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.7%
- EPSS 0.7% chance of exploitation in 30 days, percentile 53%
- public exploits
- 0
- none known
Summary
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Patch / fix
- baxter life2000_ventilator_firmware 6.08.00.00
Is CVE-2024-48966 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-24-319-01 updated cvss 10.0
Baxter Life2000 Ventilation System
-
Baxter Life2000 Ventilation System <=06.08.00.00affected: vendor fix, mitigation “Baxter plans to issue a follow-up announcement in Q2 2025 regarding the Life2000 vulnerabilities described in this disclosure.” baxter.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- yes
- Technical impact
- total
- Assessed
- last yr.
Exploit prediction
0.70% chance of exploitation in the next 30 days, higher than 53% of all scored CVEs (FIRST EPSS, 2026-11-11).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 10.0 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
Affected products
- Baxter Life2000 Ventilation System
- baxter life2000_ventilator_firmware
Classification
- Weakness
-
- CWE-306 Missing Authentication for Critical Functionality