critical 10.0

CVE-2024-48966 json

Life2000 service tools for test and calibration do not support user authentication

published last yr. modified 4mo ago also EUVD-2024-43153

cvss
10.0
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.7%
EPSS 0.7% chance of exploitation in 30 days, percentile 53%
public exploits
0
none known

Summary

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Patch / fix

  • baxter life2000_ventilator_firmware 6.08.00.00

Is CVE-2024-48966 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-24-319-01 updated cvss 10.0

Baxter Life2000 Ventilation System

  • Baxter Life2000 Ventilation System <=06.08.00.00
    affected: vendor fix, mitigation “Baxter plans to issue a follow-up announcement in Q2 2025 regarding the Life2000 vulnerabilities described in this disclosure.” baxter.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
yes
Technical impact
total
Assessed
last yr.

Exploit prediction

0.70% chance of exploitation in the next 30 days, higher than 53% of all scored CVEs (FIRST EPSS, 2026-11-11).

CVSS scores

assigned by version score vector
CNA 3.1 10.0 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Affected products

  • Baxter Life2000 Ventilation System
  • baxter life2000_ventilator_firmware

Classification

Weakness
  • CWE-306 Missing Authentication for Critical Functionality