CVE-2025-1204 json
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existin…
published last yr. modified 4mo ago also EUVD-2025-5402
- cvss
- 7.7
- version 4.0, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 42%
- public exploits
- 0
- none known
Summary
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Is CVE-2025-1204 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-25-030-01 updated cvss 7.5
Contec Health CMS8000 Patient Monitor (Update A)
-
Contec Health CMS8000 Patient Monitor Contec Health CMS8000 Patient Monitor Firmware smart3250-2.6.27-wlan2.1.7.cramfs, CMS7.820.075.08/0.74(0.75), CMS7.820.120.01/0.93(0.95)affected: mitigation “Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.” fda.gov
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- Proof of concept is public
- Automatable
- no
- Technical impact
- total
- Assessed
- last yr.
Exploit prediction
0.48% chance of exploitation in the next 30 days, higher than 42% of all scored CVEs (FIRST EPSS, 2026-11-11).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 4.0 | 7.7 | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N |
| CNA | 4.0 | 7.7 | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Contec Health CMS8000 Patient Monitor
Classification
- Weakness
-
- CWE-912 Hidden Functionality