CVE-2026-111594 json
InterSystems Corporation InterSystems IRIS for Health Cleartext Transmission of Sensitive Information
published 18d ago modified 7d ago also EUVD-2026-108523
- cvss
- 8.1
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.2%
- EPSS 0.2% chance of exploitation in 30 days, percentile 8%
- public exploits
- 0
- none known
Summary
InterSystems IRIS for Health has TLS/SSL encryption turned off by default, which may enable an attacker to view transmitted data and use that data to log into the device.
Is CVE-2026-111594 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-26-295-01 updated cvss 8.1
InterSystems Corporation InterSystems IRIS for Health
-
InterSystems Corporation InterSystems IRIS for Health 2026.2.0.221, 2026.1.0.235, 2025.3.0.226affected: mitigation “InterSystems documentation explains how to configure the server to require TLS: https://docs.intersystems.com/irislatest/csp/docbook/DocBook.UI.Page.cls?KEY=GSA_manage_superserver” docs.intersystems.com
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- no
- Technical impact
- total
- Assessed
- 7d ago
Exploit prediction
0.19% chance of exploitation in the next 30 days, higher than 8% of all scored CVEs (FIRST EPSS, 2026-11-09).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 3.1 | 8.1 | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
| CNA | 4.0 | 9.2 | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N |
| CNA | 4.0 | 9.2 | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- InterSystems IRIS for Health
Classification
- Weakness
-
- CWE-319 Cleartext Transmission of Sensitive Information
References
- docs.intersystems.com /irislatest/csp/docbook/DocBook.UI.Page.cls?KEY=ROARS_gsecuring_checklist
- docs.intersystems.com /irislatest/csp/docbook/DocBook.UI.Page.cls?KEY=GSA_manage_superserver
- wrc.intersystems.com
- cisa.gov /news-events/ics-medical-advisories/icsma-26-295-01
- github.com /cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-295-01.json
Sources
- CVE List CVE-2026-111594
- NVD CVE-2026-111594
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.