Skip to content
Fencer Intel
high 7.4

CVE-2026-67558 json

Mira Hormone Monitor, Mira Android App Authentication bypass by spoofing

published 3mo ago modified 2mo ago also EUVD-2026-57006

cvss
7.4
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.3%
EPSS 0.3% chance of exploitation in 30 days, percentile 19%
public exploits
0
none known

Summary

The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view.

Is CVE-2026-67558 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-26-223-01 updated cvss 7.4

Mira Hormone Monitor, Mira Android App

  • Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App 4.5.15.4 Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Monitor Firmware 1.7.1.47
    affected: mitigation “Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required.”

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
no
Technical impact
partial
Assessed
2mo ago

Exploit prediction

0.27% chance of exploitation in the next 30 days, higher than 19% of all scored CVEs (FIRST EPSS, 2026-11-11).

CVSS scores

assigned by version score vector
CNA 4.0 8.2 CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N
CNA 3.1 7.4 CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:L/A:N
CNA 4.0 8.2 CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:L/VA:N/SC:H/SI:L/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware
  • Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App

Classification

Weakness
  • CWE-290 Authentication Bypass by Spoofing

References

Sources

Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.