CVE-2026-68067 json
Mira Hormone Monitor, Mira Android App Weak Authentication
published 3mo ago modified 2mo ago also EUVD-2026-56982
- cvss
- 9.8
- version 3.1, CNA
- cisa kev
- Not listed
- not in CISA's catalog
- epss
- 0.5%
- EPSS 0.5% chance of exploitation in 30 days, percentile 44%
- public exploits
- 0
- none known
Summary
The login endpoint on the Mira cloud API accepts any format-valid string in the password field and returns a live active session token for the account matching the supplied email address. An attacker could use an email address to control cloud accounts and access hormone record information and account settings.
Is CVE-2026-68067 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.
Check with FencerVendor advisories
What the vendors themselves say about their products: affected, fixed or not affected, and what to do.
cisa ICSMA-26-223-01 updated cvss 9.8
Mira Hormone Monitor, Mira Android App
-
Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App 4.5.15.4 Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Monitor Firmware 1.7.1.47affected: mitigation “Users should update the Mira app to the latest version iOS v3.5.18 / Android v4.5.18. Firmware v01.07.01.53 is updated via the app when the device is connected. No additional action is required.”
From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf
CISA SSVC
- Exploitation
- none
- Automatable
- yes
- Technical impact
- total
- Assessed
- 2mo ago
Exploit prediction
0.51% chance of exploitation in the next 30 days, higher than 44% of all scored CVEs (FIRST EPSS, 2026-11-11).
CVSS scores
| assigned by | version | score | vector |
|---|---|---|---|
| CNA | 4.0 | 9.3 | CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N |
| CNA | 3.1 | 9.8 | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| CNA | 4.0 | 9.3 | CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X |
Affected products
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App
Classification
- Weakness
-
- CWE-1390 Weak Authentication
References
Sources
- CVE List CVE-2026-68067
- NVD CVE-2026-68067
Fencer Intel keeps one entry per vulnerability: the CVE id wins, then GHSA, and records that alias each other are merged, except that two CVEs always keep separate entries. The same data is available as JSON.