CVE-2026-7251 json

Eppendorf BioFlo 320 Use of hard-coded password

published 5mo ago modified 3mo ago also EUVD-2026-31912

cvss
9.8
version 3.1, CNA
cisa kev
Not listed
not in CISA's catalog
epss
0.5%
EPSS 0.5% chance of exploitation in 30 days, percentile 43%
public exploits
0
none known

Summary

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Is CVE-2026-7251 in your stack? Fencer shows which of your repositories are affected as soon as an advisory lands, and opens the fix as a pull request.

Check with Fencer

Vendor advisories

What the vendors themselves say about their products: affected, fixed or not affected, and what to do.

cisa ICSMA-26-146-01 updated cvss 9.8

Eppendorf BioFlo 320

  • Eppendorf BioFlo 320 Bioreactor
    affected: mitigation “Eppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads.” eppendorf.com

From CISA ICS and IT advisories (CSAF, https://github.com/cisagov/CSAF), a U.S. Government work; changed: reduced to products, statuses and remediations. CISA notification · csaf

CISA SSVC

Exploitation
none
Automatable
yes
Technical impact
total
Assessed
5mo ago

Exploit prediction

0.50% chance of exploitation in the next 30 days, higher than 43% of all scored CVEs (FIRST EPSS, 2026-11-11).

CVSS scores

assigned by version score vector
CNA 4.0 9.3 CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
CNA 3.1 9.8 CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CNA 4.0 9.3 CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Affected products

  • Eppendorf BioFlo 320

Classification

Weakness
  • CWE-259 Use of Hard-coded Password

References